Cyclic Fatigue and Torsional Failure of EdgeTaper Platinum Endodontic Files at Simulated Body Temperature

Jun 12, 2019 | Studies

Ahmed Jamleh, BDS, MSc, PhD, Amjad Alghaihab, DMD,
Abdulmohsen Alfadley, BDS, MSc, FRCD(C), Hussam Alfawaz, BDS, MS, Abdullah Alqedairi, BDS, MS, and Khalid Alfouzan, BDS, MSEd

Introduction: The aim of this in vitro study was to compare the cyclic fatigue resistance and torsional resis-tance of EdgeTaper Platinum (ETP) with those of Pro-Taper Gold (PTG) at simulated body temperature.

 

Methods: ETP and PTG files with #25 tip size were selected. Cyclic fatigue tests were performed in a stain-less steel artificial canal until file fracture occurred. The time to fracture was recorded. For the torsional resis-tance testing, the apical 5 mm of the file was firmly secured with acrylic resin, and the assembly was fixed over torque gauge device. A uniform torsional stress was applied with continuous rotation motion until frac-ture occurred. The torque at failure was recorded. Both experiments were conducted at 35 1 C. Fractured sur-faces were tested via scanning electron microscopy. Sta-tistical analysis was performed using independent Student t test at 5% significance level.

Results: The ETP had superior cyclic fatigue resistance than PTG (P < .01). However, it exhibited lower torsional resis-tance (P < .05). The ETP and PTG showed typical fea-tures of cyclic and torsional fatigue behavior under scanning electron microscopy.

Conclusions: The ETP exhibited superior resistance to cyclic fatigue but failed to show any improvement in the torsional resistance compared with PTG. (J Endod 2019;45:611–614)

Related Posts

Evaluation of the Amount of Root Canal Dentin Removed and Apical Transportation Occurrence after Instrumentation with ProTaper Next, OneShape, and EdgeFile Rotary Systems

Abstract Introduction The aim of this study was to evaluate the amount of root canal dentin removed and apical transportation occurrence after instrumentation of mesiobuccal canals of maxillary molars with ProTaper Next (PTN [Dentsply Maillefer, Ballaigues,...

Differences in cyclic fatigue lifespan: WaveOne Gold vs EdgeOne Fire.

Differences in cyclic fatigue lifespan between two different heat treated NiTi endodontic rotary instruments: WaveOne Gold vs EdgeOne Fire. Gambarini G, Galli M, Di Nardo D, Seracchiani M, Donfrancesco O, Testarelli L. J Clin Exp Dent. 2019 July 1;11(7):e609-e613. *...

Mechanical Properties of Various Glide Path Instruments

Highlights •Cyclic resistance, torsional resistance, and bending stiffness of glide path files made of different alloys were compared using custom devices. •EdgeGlidePath made from heat-treated wire showed higher cyclic resistance than ProGlider and One G. •One G made...

Transportation Study

For printable PDF version click here: Cone-beam Computed Tomographic Analysis of Canal Transportation and Centering Ability of Single-file Systems

AAE18 Research Evaluates EdgeEndo® Files

Abstracts of Research These abstracts are Oral and Poster Research that will be presented at AAE18 . Abstracts evaluated EdgeEndo® files, with the majority of studies evaluating the EdgeTaper Platinum™ vs ProTaper® Gold. Abstracts appear as they were submitted by the...

Low Cost Endodontic Files

Low Cost Endodontic Files A publication of CR Foundation ® Reprinted June 2017, with permission, from Vol. 10 Issue 6, Pages 1,4,5 Gordon’s Clinical Observations: Endodontic treatment is a nearly daily procedure for most general dentists, and the relatively low...